USB charging/download cable specifications
Input/Output
5 V DC, 1A
Type
USB A to USB micro B
Length
0.91 meters
Power supply/charger specifications
Class
II
Input
AC Input 100–240 Vac, 50/60Hz, 0.2A, 0.2A rms at 100 Vac
DC output
5V DC, 1A (5.0 Watts)
Electromagnetic immunity and emissions declaration and guidance
The transmitter and receiver are intended for use in the electromagnetic environment specified in the next table. The customer
or the user of the transmitter should ensure that it is used in such an environment.
Immunity test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Magnetic Field (50Hz)
IEC 61000-4-8
Electrical Fast Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips and Interruptions
IEC 61000-4-11
IEC 60601-1-11
Conducted Fields Disturbance
IEC 61000-4-6
Radiated Fields Disturbance
IEC 61000-4-3
Radiated and Conducted Fields
Aircraft use
Transmitter compliance level Receiver compliance level
± 8 kV Contact
± 15 kV Air
30 A/m
N/A
N/A
N/A
N/A
10 V/m
at 80 MHz to 2700 MHz (AM Modulation)
Meets RTCA /DO-160 edition G Section 20, Category T
F • Technical information |
77
± 2 kV for power supply lines
± 0.5 kV, ± 1 kV line(s) to line(s)
0% 230V for 1 cycle
0% 230V for 0.5 cycle at 8 phase angles
70% 230V (30% dip in 230V) for 25 cycles
0% 230V for 250 cycles
6 Vrms
150 kHz to 80 MHz