Set-Up
It is recommended that the calibration code
be changed after calibration in order to
prevent unauthorized tampering. Authorized
calibration personnel should record and
store the new calibration codes. See
Calibration Code Section and Entering a
PIN Code, page 18.
8.
Select VIEW LOG.
F
. 20
IG
NOTE: The value in the K-F column for the
most recent record date is the current
calibration factor (Kcurrent). For new meters
that have not been previously calibrated
Kcurrent = 1730.
Current calibration factor
F
. 21
IG
9.
Calculate the new K-Factor using the
following equation:
NOTE: The unit of measurement for both
volumes must be the same.
VOLUME DISPENSED, see Step 3, page 15.
VOLUME DISPLAYED ON METER, see Step
4, page 15.
16
Kcurrent, see Steps 5-8, beginning on page
15.
Example:
Kcurrent = 1710
Volume displayed on metered dispense
valve = 4.970 liters
Volume dispensed = 5 liters
(1710) x (4.970 liters)
Knew
=
(5 liters)
Knew
Round to the nearest whole number: 1699.74
= 1700.
10. Enter the new calibration factor (Knew)
into the meter.
11. Select SET-UP.
F
. 22
IG
12. Select CALIBRATE.
F
. 23
IG
13. Enter CALIBRATION CODE (PIN).
=
1699.74
3A7397E