Err 220
ROM checksum error; the data in the internal
ROM of the balance processor (AOC) are
incorrect.
Err 230
RAM read/write error; the data in the internal
RAM of the balance processor AOC is incorrect,
or not possible.
Err 237
EEPROM checksum error in the linearity range; Overwrite the incorrect data record using the
the balance has not yet been linearized, or the MC1 Server 4.4. Then proceed as directed in
data in the internal EEPROM of the balance
processor AOC are incorrect.
Err 239
EEPROM checksum error in the linearity weight Follow the procedure for "Overwriting the Internal
range; the stored factor for the linearity weight
is not within the permissible tolerance range or
is incorrect.
Err 241
EEPROM checksum error in the permanent area; Overwrite the incorrect data record using the
the data for the balance operating menu in the MC1Server 4.4. If the error remains, then proceed
internal EEPROM of the balance processor
AOC are incorrect.
Err 243
EEPROM checksum error in the menu range;
the inalterable data in the internal EEPROM
of the balance processor AOC are incorrect.
Err 245
EEPROM checksum error in the calibration/
adjustment range (zero point); the balance has
not yet been calibrated or the data in the internal
EEPROM of the balance processor AOC are
incorrect.
Err 247
EEPROM checksum error in the adjustment/
calibration range (span); the balance has not
yet been calibrated or the data in the internal
EEPROM of the balance processor AOC are
incorrect.
Err 249
EEPROM checksum error in the calibration
weight range; the factor stored for the internal
weight calibration exceeds the tolerance range,
or is incorrect.
L
The weight on the pan is below the weighing
capacity; the pan is not on the balance,
the balance was incorrectly adjusted
or the preload was incorrectly adjusted.
H
The weight on the pan exceeds the weighing
capacity; the weight is too heavy; the balance remains, perform "External Calibration/Adjustment"
was incorrectly adjusted;or the preload was
incorrectly adjusted.
26
Follow the procedure in "Exchanging the Main PCB."
Follow the procedure in "Exchanging the Main PCB."
"Adjusting the Linearity" and "External Calibration."
Linearity Weight."
with "Exchanging the Main PCB."
With the menu access switch unlocked, access the
menu (display will flash). Exit the menu and store (press
the tare key for more than 2 sec.) => The menu is now
reset. Checksums are corrected.
Perform "External Calibration/Adjustment."
Perform "External Calibration/Adjustment."
Follow the procedure for "Overwriting the Internal
Calibration Weight."
Make sure there was no operating error. If the error
remains, perform "External Calibration/Adjustment"
or follow the procedure for "Adjusting the Preload."
Make sure there was no operating error. If the error
or follow the procedure for "Adjusting the Preload."

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Basicplus and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Bp210dBp300s

Table of Contents

Save PDF