scale value set.
FILTER VALUE
0
1
2
3
4 (default)
5
6
7
8
9

RESETTABLE WEIGHT SETTING FOR SMALL WEIGHT CHANGES

 (from 0 to max full scale; default: 300; considered decimals: 300 – 30.0 – 3.00 – 0.300): this
parameter indicates the maximum weight value resettable by external contact, keypad or serial
protocol.
 (from 0 to max 20% of full scale; default: 0): If at switch-on the weight value is lower than
the value set in this parameter and does not exceed the  value, the weight is reset. To
disable this function, set 0.
 (from 1 to 5, default: ): When the weight value is stable and, after a second, it
deviates from zero by a figure in divisions smaller or equal to the figure in divisions set in this
parameter, the weight is set to zero. To disable this function, set .
Example: if the parameter  is set to 5 and   is set to 2, the weight will be automatically
set to zero for variations smaller than or equal to 10 ( x  ).
Response times
[ms]
12
150
260
425
850
1700
2500
4000
6000
7000

ZERO PARAMETERS

AUTOMATIC ZERO SETTING AT POWER-ON

ZERO TRACKING

Display and serial port refresh
- 23 -
frequency
[Hz]
300
100
50
25
12.5
12.5
12.5
10
10
5