Setpoint can be used to signal that totalized partial weight and total weight have been
reached.
WTAB: setpoint are always enabled.
WDESK – WINOX: proceed as follow to enable setpoint:
-(default): setpoint disabled;
-: setpoint enabled;
Setpoint 1 and 2 switch respectively based on totalized partial weight and total weight, the others
always switch on weight.
After enabling the setpoint, key operations to access menus vary as follows:
- Press MENU to access setpoint settings;
- Press START for 3 seconds to change operating mode.
the load cell is not connected or is incorrectly connected; the load cell signal exceeds
:
39 mV; the conversion electronics (AD converter) is malfunctioning; the load cell is a 4-
wire and there are no jumpers between EX- and REF- and between EX+ and REF+.
:
the weight display exceeds 110% of the full scale.
internal instrument converter failure; check load cell connections, if necessary contact
:
technical assistance.
: the weight exceeds the maximum capacity by 9 divisions.
:
maximum displayable value exceeded (value higher than 999999 or lower than -999999).
: weight too high: zero setting not possible.
: this message appears in the sample weight setting, in real calibration, after the fifth
sample weight value has been entered.
:
the value set for the parameter is beyond the permitted values; press ESC to quit the
setting mode leaving the previous value unchanged. Examples: a number of decimals is
selected for full scale which exceeds the instrument's display potential; value above the
maximum setting value; the weight value set in sample weight verification does not match
the detected mV increase; the analog output correction goes beyond the permitted limits.
:
lock active on menu item, keypad or display.
: It's not possible to display properly the number because is greater than 999999 or less
than -999999.
: buffer battery low, loss of date and time of Real-Time Clock. Confirm with ENTER to
continue; leave the instrument on for at least 12 hours to charge the battery, if the alarm
persists contact technical assistance.
an incorrect date has been detected: go into the related menu to check and correct it.
:

ENABLING SETPOINT