Abbott Ultiri 1015082 Operation Manual page 37

Table of Contents

Advertisement

Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Immunity to Proximity
Fields from RF
wireless
communications
equipment
IEC 60601-1-2
(Clause 8.10)
Guidance and Manufacturer's Declaration (continued)
IEC 60601 Test Level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
385–5785 MHz
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects, and
people.
a
Field strengths from fixed Transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast,
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF Transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the device is used exceeds the applicable RF compliance
level above, the device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
Ultiri™ Measurement System Operation Manual – 600327583 Rev. A
Compliance Level
3 Vrms
[V
= 3]
1
3 V/m
[E
= 3]
1
9–28 V/m
Specifications
Electromagnetic Environment -
Guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the device, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the Transmitter.
Field strengths from fixed RF
Transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Per IEC 60601-1-2 (per Table 9 of
Standard).
a
] V/m.
1
33

Advertisement

Table of Contents
loading