Wireless Communication

The Ultiri Measurement System utilizes radios to enable the wireless connectivity
of the Wi-Box™ AO Transmitter, the PressureWire X Guidewire, and the
tableside controller (TSC), as shown in this figure. The details of each wireless
component are covered in the table below.
Wireless
Component
PressureWire X
Transmits distal pressure value to Ultiri
Guidewire
Measurement System.
Wi-Box AO
Transmits aortic pressure value to Ultiri
Transmitter
Measurement System; wired connection
to hemodynamic monitoring system.
TSC
Transmits control commands to Ultiri
Measurement System.
Note:
If the TSC fails to connect at any time during the procedure, use the
touchscreen to continue the procedure until the TSC can be
reconnected by following the guidance in "Troubleshooting" on page 27.
If the PressureWire X Guidewire or the Wi-Box AO Transmitter fails to
connect at any time during the procedure, a "No Sensor" error message
is displayed. Reconnect the component by following the guidance in
"Troubleshooting Physiology Procedures" on page 10.
Ultiri™ Measurement System Operation Manual – 600327583 Rev. A
Ultiri Measurement System
Function
Safety Information
Operating Distance
Up to 4 m (13.1 ft)
Up to 4 m (13.1 ft)
Up to 3 m (9.8 ft)