Electrical Shock Hazard - Do not remove Ultiri Measurement System
covers. To avoid electrical shock, use only the power supply and power
cord supplied with the system. Refer to "Setting Up the System" and
"Safety Information" for electrical safety information.
The main power remains switched on when the system is in Standby
mode. Avoid direct or indirect (e.g., via the operator) conductive
connection between other electrical equipment and the Ultiri
Measurement System. Conductive connection may cause leakage
currents to induce ventricular fibrillation. High frequency surgical
equipment must not be used on a patient at the same time as
PressureWire X Guidewire and the Ultiri Measurement System.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than
30 cm (12 in) to any part of the Ultiri Measurement System, including
cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Do not use the Ultiri Measurement System if it has been dropped, or in
any other way exposed to mechanical or electrical damage, or if liquids
are suspected to have penetrated the casing or the power supply unit.
This can cause the user to be exposed to electric shock or loss of
system function. Contact Abbott Technical Service for further
instructions.
Do not sterilize the Ultiri Measurement System or any of its parts. Do not
use this system or any of its parts if it has been sterilized.
The Ultiri Measurement System contains a lithium battery for the system
real-time clock. Danger of explosion. Battery is not intended to be
replaced.
To prevent possible patient or operator injury and damage to the system,
make sure the equipment is securely mounted.
Make sure the unit is dry before connecting it to the power supply.
The PressureWire X Guidewire is sterilized by ethylene oxide and is
intended for one-time use only. Non-pyrogenic. Do not use if the
package is opened or damaged. Do not reuse or re-sterilize. Any
attempt to reuse or re-sterilize the PressureWire X Guidewire may
compromise the structural integrity of this device. Adverse effects of
using a non-sterile or re-sterilized guide wire may include, but are not
limited to: local and / or systemic infection, mechanical damage,
inaccurate results.
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such
use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally. If abnormal
performance is observed, it is necessary to increase distance between
equipment.
Ultiri™ Measurement System Operation Manual – 600327583 Rev. A
Warnings, Precautions, and Limitations
15