Electromagnetic Immunity

The device is intended for use in the electromagnetic environment specified
below. The customer or user of the device should ensure that it is used in such
an environment.
Guidance and Manufacturer's Declaration
IEC 60601 Test Level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input /
output lines
±1 kV differential
mode
0% U
; 0.5 cycle
T
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% U
; 1 cycle
T
and
70% U
; 25 / 30
T
cycles
Single phase: at 0°
0% U
; for 5 sec @
T
60 Hz (300 cycles)
0% U
; 250/300
T
cycles
30 A / m
Note:
is the AC mains voltage prior to application of the test level.
U
T
Compliance Level
±8 kV contact
±15 kV air
±2 kV
±1 kV
±1 kV
100% dropout in VNOM
for 0.5 cycle at listed
phase angles
100% dropout in VNOM
for 1 cycle at 0°
30% dropout in VNOM
for 25 / 30 cycles at 0°
100% dropout in VNOM
for 5 sec
100% interrupt in VNOM
for 250/300 cycles
30 A / m
Electromagnetic Environment -
Guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
device requires continued
operation during power mains
interruptions, it is recommended
that the device be powered from
an uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.