Setting Up the System

Ensure that the power cord and all other connections to the Ultiri™
Measurement System are routed to prevent a tripping hazard.
Ensure that the main power switch and power plug can be accessed at any time
during the procedure.
Position the cables to minimize the chance of disconnection.
Mount the TSC onto the procedure table rail in a location that is comfortable for
the physician to use during the procedure. Drape the TSC with a protective
sterile barrier to maintain sterile technique.
Ensure that the TSC is in a location that minimizes exposure to fluids.

Attaching and Removing the TSC (Table Rail)

Attach
Ensure that the clamp is fully open.
1.
Place the clamp side of the component onto the rail and continue to hold.
2.
Gently push down on the grooved end of the clamp to secure the component
3.
onto the rail.
Gently test the component to ensure that it is securely fastened to the rail.
4.
Note: If a component is moved to a different table, a shim may be
required to allow it to clamp properly.
5.
Plug the USB cable from the TSC power supply into the bottom of the TSC.
WARNING: Do not allow the TSC power supply to get wet.
Remove
Unplug the USB cable from the TSC power supply from the bottom of the TSC.
1.
Grasp the component from the bottom in one hand (supporting the weight) and
2.
lift the grooved end of the clamp to release from the rail.
3.
Holding the clamp open, gently pull the component from the rail. To avoid
damaging cables, components, or connections that may expose the user to
harm, do not drop the component.