Emissions Test
RF Emissions, CISPR 11
RF Emissions, CISPR 11
Harmonic Emissions,
IEC 61000-3-2
Voltage Fluctuations /
Flicker Emissions,
IEC 61000-3-3

Radio

Parameter / Category
Standards
Data Rate
Operating Frequency
Modulation

Electromagnetic Emissions

The Ultiri™ Measurement System is intended for use in the electromagnetic
environment specified below. The customer or user of the Ultiri Measurement
System should ensure that it is used in such an environment.
Guidance and Manufacturer's Declaration
Compliance
Group 1
Class A
Class A
Complies
Note: The emission characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required), this equipment
might not offer adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or
reorienting the equipment.
Ultiri™ Measurement System Operation Manual – 600327583 Rev. A
Specification
IEEE 802.11ac/a/b/g/n (2T2R), Bluetooth
V4.2, V4.0 LE, V3.0+HS, V2.1+EDR
802.11b: 11 Mbps/802.11a/g: 54 Mbps/802.11n:
MCS0~15/802.11ac: MCS0~9
Bluetooth
: 1 Mbps, 2 Mbps, and up to 3 Mbps
®
EDR
IEEE 802.11 ac/a/b/g/n ISM Band,
2.412 GHz~2.484 GHz, 5.150 GHz~5.850 GHz.
Subject to local regulations
802.11b: DSSS (DBPSK, DQPSK, CCK)
802.11a/g: OFDM (BPSK, QPSK, 16-QAM,
64-QAM)
802.11n: OFDM (BPSK, QPSK, 16-QAM,
64-QAM)
802.11ac: OFDM (BPSK, QPSK, 16-QAM,
64-QAM, 256-QAM)
Electromagnetic Environment Guidance
The Ultiri Measurement System uses RF
energy only for its internal function.
Therefore, its RF emissions are very low and
not likely to cause any interference with
nearby electronic equipment.
The emission characteristics of the Ultiri
Measurement System are suitable for use in
industrial areas and hospitals (CISPR 11
class A). The System should not be used in
a residential (domestic) environment (CISPR
11 class B is required in support of domestic
use).
Specifications
V5.0,
®
31