2 Safety notes

2.4 Measures and duties of the operator

14 / 116
2.4 Measures and duties of the operator
Electrical safety
Switch the system off every time before disconnecting it from
u
the power supply or if you are not going to use the device for
any length of time. Also disconnect the system from the power
supply before cleaning surfaces or accessories.
Insert the power plug only into a power outlet that has a
u
functioning protective earth connection.
Use only cables and plugs which are in perfect working
u
condition.
Connect the laser console only to a power supply that corre-
u
sponds to the values specified on the rating plate.
Do not use multiple sockets!
u
Do not use extension cables!
u
Do not touch the system when your body is charged electro-
u
statically and the system is not grounded.
Observe the instructions regarding electromagnetic compati-
u
bility (EMC).
Set the system up so that the power cable can be disconnected
u
from the power supply quickly and without any supplementary
means.
Connect the laser console using the power cable intended for
u
use with the laser console. If the laser console is mounted on
an IT 1060.i / IT 760.i instrument table, it will be powered
through this table.
Perform the electrical installation in conformance to
u
IEC 60364-7-710 or the applicable national regulations. This
includes the integration of a ground fault circuit interrupter
(GFCI).
Do not touch the patient and the connections of the system
u
simultaneously.
Live parts are accessible inside the system. If you remove the
housing, you are exposed to the risk of an electric shock.
Never open the system!
u
Ambient conditions
Make sure that the installation requirements and the operation
u
of the system meet the following requirements:
• Low vibration
• Clean environment
• Avoid extreme mechanical loads
Do not operate the supplied devices, provided that they are
u
energetically operated,
Documentation set
VISULAS green
000000-2096-101-DokS-GB-181218