External Weight Calibration - Excell SI-132 Service Manual

High resolution weighing scale
Hide thumbs Also See for SI-132:
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD

2-2 External Weight Calibration

Press TARE key,
When the right digit is flashing,
put the platter on the scale and
press TARE key while there is
no load on platter.
After stable
According to the calibration
weight to put the weight on the
platter and press TARE key.
The scale is managing the data
of full capacity. When it
becomes stable, the beeper
makes 3 beep sounds and the
display flickers. Then take off
finish setting
the weigh to
Conditions for external weight calibration:
The calibration weight value placed on the platter must be over 100e, and the standard
deviation of the weight must be within 10%.
Approval model cannot access
02005302
kg
.
kg
kg
.
21
the display shows the calibration
value (the maximum capacity of 1
UNITS setting) and UNITS. The
right digit keeps flashing.
The calibration value could be
changed.
The scale is managing the data of
zero point. When it becomes stable,
the display shows the calibration
weight.
ON/ZERO key ⇒ Upward key (0~9 digit entry)
UNITS key ⇒ Downward key (0~9 digit entry)
TARE key ⇒ Move cursor rightward
NET/GROSS key ⇒ Move cursor leftward