Table of Contents PRECAUTIONS FOR USE ....................2 QUICK SETUP CALIBRATION .................... 4 CHAPTER 1 INTRODUCTION .................... 5 1-1 F ..................5 EATURES AND PECIFICATIONS 1-2 P ......................5 OWER UPPLY 1-3 P ................7 ANEL AND EYBOARD NTRODUCTION 1-4 O ....................
Page 3
EXCELL PRECISION CO., LTD. 3-4-2 Approval Configuration ............36 3-4-3 Buzzer Device Setting ............. 37 3-4-4 Initial Zero Setting ..............38 3-4-5 Tare Function Setting ............... 39 3-4-6 Software Seal Function Setting ..........40 3-5 L ..............41 OCAL RAVITY ETTING APPENDIX 1 7-SEGMENT DISPLAY CHARACTERS............
EXCELL PRECISION CO., LTD. Thank you for purchasing EXCELL waterproof weighing scale. In order to use the scale properly, please read this instruction carefully before use. If you have a problem concerning the scale, please contact your supplier. Preparing to Use the Scale 1.
Page 5
EXCELL PRECISION CO., LTD. a) Battery is installed but the contact is poor, please check whether the battery is installed properly. b) Please check whether the polarity of the battery is reversed. When the polarity is reversed, the “Charging” indicator will flash after charging for a period of time.
EXCELL PRECISION CO., LTD. Quick Setup Calibration This page is to quickly initiate the scale, for the other functions configuration, you can refer the chapters below. Instructions: Step 1 Power off the scale, find the switch SWA1 on the bottom of machine.
EXCELL PRECISION CO., LTD. Chapter 1 Introduction 1-1 Features and Specifications Features: Up to 1/7,500 display resolution available. IP68 and IP 69 waterproof and dustproof design Food grade SUS304 stainless steel housing and weighing pan. Futuristic streamlined design, lightweight and portable, integrated ...
Page 8
AC cable first and turn on scale, and then check the number of bars under battery symbol for battery level. Notice for SUBe AirCharge for Wireless Charging a. While the scale is charging, the spacing between charging plate and scale shall be within 2mm.
EXCELL PRECISION CO., LTD. 1-3 Panel and Keyboard Introduction Zero point indication : Upper limit : Value between HI and LO Stable The weight is stable : Lower limit Net weight indication : Battery level indication (only for newer scale) Low power indication Charging indication.
EXCELL PRECISION CO., LTD. Put a container on the weigh pan and after the weight is stable, press the to zero the weight of the container. The screen displays the “Net” icon. Put the goods in the container, the screen displays the net weight value of the goods.
EXCELL PRECISION CO., LTD. INTERNAL VALUE MODE (must hook up Load Cell to test) Press to enter, and the display shows internal value Please check whether the internal value has changed obviously with weight changing. Please check the backlight.
EXCELL PRECISION CO., LTD. 1-6 Weight Unit 1 g = 0.001 kg 1 g = 1 g 1 g = 0.002204623 lb 1 g = 0.03527396 oz Chapter 2 Advanced Functions In the weighing mode, press the keys at the same time to enter the Advanced Function setting mode.
EXCELL PRECISION CO., LTD. 2-1-1 FnC 01 Power Saving Function Setting Power Saving Function Setting the setting last time. Default setting = oFF Displays Press key to switch On = power saving on Off = power saving off Press When the weight is over 10d, the display shows normal.
EXCELL PRECISION CO., LTD. 2-1-3 FnC 03 Hi/Lo/OK Function Setting When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled. This function is available in all unit modes. In one specific unit mode, enter FnC 03 to set the Hi/Lo/OK values.
EXCELL PRECISION CO., LTD. 2-1-5 FnC 05 Noise Filter Setting Only available for non-approval models. (CFn 02 = 0) Returning to the zero point setting key to key in the parameters or zero point. Default setting = 0 Press When the weight on the scale is over 1/3 full capacity, the function is on.
EXCELL PRECISION CO., LTD. 2-1-7 FnC 09 Unstable Tare Only available for non-approval models. (CFn 02 = 0) Unstable Tare setting the last saved parameter setting Displays key to key in the parameters for second Press weighing unit Default setting = 0 0 ...
EXCELL PRECISION CO., LTD. 2-2 02 EC Weight Calibration Press the calibration weight value (the Displays maximum capacity of 1 unit) and the unit. The digit on the right side keeps flashing. When the digit on the right flashes, ensure that the weigh The scale is calculating the internal value of pan is empty.
EXCELL PRECISION CO., LTD. Chapter 3 Service Mode Access Set the switch SWA1 on the bottom of machine to the ADJ position. Switch the machine on. The display shows 01 CSP. When finished, set the switch SWA1 back to the LOCK position.
EXCELL PRECISION CO., LTD. 3-1 Capacity Setup Press Input a parameter “00”-“03” After inputting “00”, press key to return to the previous step. (Exiting from capacity setup) Return to the Upper Level Weight Unit Setting Customized Weight Unit Setting ...
EXCELL PRECISION CO., LTD. 3-1-1 Weight Unit Setting The weight UNITS ”kg, g, lb” have no resolution limit. But the weight UNITS ”oz, GN, dwt, and ct” are only available on indicators with less 1/10,000 external resolution. The display shows current setting Press is max capacity with 0.01kg/division...
Page 21
EXCELL PRECISION CO., LTD. NOTE 1 (a) (b) (c) (d) (e) (f) (a) The first weight UNITS (only “kg”, “g”, or “lb” are available to choose from. Please select one of the parameters 0, 1, or 2) (b) The second weight UNITS (select one of the parameters described below) (c) ...
Page 22
EXCELL PRECISION CO., LTD. 2 Segment range: multi_interval multi_range 3 Segment range: multi_interval multi_range (q) Division multiplication factor: (Only one Weight UNIT Model is available) 0 no factor 1 base value X 10 Division Table for various m and q values:...
EXCELL PRECISION CO., LTD. 3-1-2 Customized Capacity Setting Resolution has no high limit, but over 1/10000 will result in error. The display shows current setting is max capacity of 6 lb with 0.001 Press lb/division for 2 seconds. The display shows the number of weighing...
Page 24
EXCELL PRECISION CO., LTD. Cont Press The display shows current setting is max capacity of 12 lb with 0.005 lb/division for 2 seconds. apacity +1d and UNITS weight for Enter 006000 6000 g 2 seconds. Refer to “Note 5”...
Page 25
EXCELL PRECISION CO., LTD. NOTE 4 (a) (b) (a) The number of the weight UNITS (Max: 5, key in 1 ~ 5) (b) The weight UNITS for weight calibration (choose from “kg”, “g”, “lb”, key in 0, 1, or 2) Parameter description: 0 ...
Page 26
EXCELL PRECISION CO., LTD. B. Choose “lb” (decimal notation system) as the weight UNITS for the weight calibration The first weight UNITS setting: 12.000 lb / 0.005 lb (the maximum Capacity / division) Enter the maximum capacity 012000 at (c) ~ (h) ...
Page 27
EXCELL PRECISION CO., LTD. NOTE 6 (o) (p) ( q ) (r ) (s ) (t) (o) Minimum division setting Parameter description: Decimal system: Hexadecimal notation system: Input 1, 2, or 5 as the Input 1, 2, 4, or 8 as the...
Page 28
EXCELL PRECISION CO., LTD. Customized Capacity Setting- How to Revise To revise customized setting, please follow the steps below Press The display shows current setting is max capacity of 12 lb with 0.005 lb/division for 2 seconds. According to Display shows parameter settings NOTE 4, re-key in from last time.
EXCELL PRECISION CO., LTD. 3-1-3 Multi-range Setting Select the desired unit for setting. All the units which set at Press could be selected to set the interval point. Refer to Note 1 on page 24 for unit Input the unit parameter setting.
EXCELL PRECISION CO., LTD. 3-2 Linearity Calibration After linearity calibration, it is necessary to re-calibrate the weight. The last point of Weight calibration should be included in linearity calibration. linearity calibration with plastic platter or The first point of linearity calibration with aluminum support.
Page 31
EXCELL PRECISION CO., LTD. Weight factor The weight factor is a single hexadecimal number which represents the value of the next weight compared to the size of the first weight applied to the scale. Assume that the weight factor of the first counterweight is n...
EXCELL PRECISION CO., LTD. 3-3-1 Weight Calibration Setting Press key, the display shows the calibration value (the maximum capacity of 1 UNITS setting) and Press UNITS. The right digit keeps flashing. The calibration value can be changed. According to the flashing number, put on the plate and press key.
EXCELL PRECISION CO., LTD. 3-3-2 Manufacturing Location Gravity Setting The gravity value should be among the value of Equator and Polar. Equator gravity G = 9.7803184558 m/sec Polar gravity G = 9.8321772792 m/sec Taipei ≒ 9.78914 m/sec Shanghai ≒ 9.79423 m/sec...
EXCELL PRECISION CO., LTD. 3-3-3 Seal Verification Test Press key to enter the Seal V-Test. Display extra digit to perform verification test here. key Upward key (from 0 to 9) Press key to exit the Seal V-Test. key Downward key (from 9 to 0) key ...
EXCELL PRECISION CO., LTD. 3-4 Function Setting Press Input a parameter 00~06 To exit enter “00” and key Upward key (from 0 to 9) Press to go back to the upper level key Downward key (from 9 to 0) key ...
EXCELL PRECISION CO., LTD. 3-4-1 Environment Parameters If parameters are changed in , then will be revised automatically. Zero-return The display shows the last saved setting Press Zero-return key to enter Default setting = 0 0 show all 5 ...
EXCELL PRECISION CO., LTD. Buzzer Device Setting 3-4-3 Press Buzzer type fitted The screen shows last saved parameter setting key to key in the parameter Buzzer type fitted key to key in the parameter 0 Pulse signal (DC type) Press 1 ...
EXCELL PRECISION CO., LTD. 3-4-4 Initial Zero Setting Zero range setting at switch on the last saved setting Displays Press Initial Zero Setting key to to key in the parameter key in parameter Default setting = 0 (OIML or NTEP approval model) Default setting = 9 (Non-approval model) 0 ...
EXCELL PRECISION CO., LTD. 3-4-5 Tare Function Setting FRQ Tare Count Settings The display shows last saved parameter settings Press FRQ Tare Count Settings The display shows last saved parameter settings key to Please Use key to select enter parameters.
EXCELL PRECISION CO., LTD. Here are the default CFn 05 settings for each approval settings of CFn 02. Parameters are fixed and cannot be changed if CFn 02 sets to 2 and 3 (Sri Lanka approval), or sets to 4 (Brazil business).
EXCELL PRECISION CO., LTD. To read software seal event counter: When switching on the scale, the software seal event counter will show up as C XXX, where XXX is 0~999. Note: SWA1 of the scale is in LOCK position.