2-4 Weight Calibration

In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to
enter the Advanced Function setting mode. The LCD shows
the NET/GROSS or UNITS key to select
mode.
ACTIONS
Press TARE key
When the digit on the right
flashes, ensure that the
weigh pan is empty. Then
Press the TARE
key.
After stabilization
Place the displayed weight
on the weigh pan. Press the
TARE key. The scale
calculates the full capacity
internal value and after
stabilization, the buzzer
beeps 3 times. Remove the
weight from the weigh pan
and the procedure
is completed.
 Weight calibration conditions:
The calibration weight value placed on the weight pan must be over 100e, and must
be within the 90% and 110% of the full weight
02026010
DISPLAY
kg
.
kg
kg
.
20
to enter the weight calibration
NOTE
The LCD displays the calibration
weight value (the maximum capacity
st
of 1
unit) and the unit. The digit on
the right side keeps flashing.
The scale is calculating the internal
value of the zero point. After
stabilization, the LCD shows the
calibration weight value.
ON/ZERO key  Upward key (from 0 to 9)
UNITS key  Downward key (from 9 to 0)
TARE key  Move the cursor to right
NET/GROSS key  Move the cursor to left.
and use
ZSME300000369