Approval Configuration - Excell SI-132 Service Manual

High resolution weighing scale
Hide thumbs Also See for SI-132:
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD

3-4-2 Approval Configuration

Press
key
TARE
Use
or
ON/ZERO
to input a parameter
Press
key
TARE
Input the parameter "00"
Press
key
TARE
to complete the setting
Brazil approved model: In the counting mode, the indicator is not able to calculate the count
value if the UNITS weight is less than 0.1d (d=division). It will be showing "------".
02005302
key
UNITS
Approval configuration
Display shows last setting
Approval configuration
Use
or
ON/ZERO
0 ⇒ Non-approval models
1 ⇒ OIML or NTEP approved model,
2 ⇒ Sri Lanka approved model, zero can
be accepted within a range of the zero
calibrated point +/- 3%.
3 ⇒ Sri Lanka approved model, zero can
be accepted within a range of the zero
calibrated point +/- 4%.
4 ⇒ Brazil approved model.
5 ⇒ Approved model. In the tare mode, the
gross weight will be displayed after
pressing the
seconds, the net value will be
displayed automatically.
External weight calibration is not available
for all approved models.
ON/ZERO key ⇒ Upward key (0~9 digit entry)
UNITS key ⇒ Downward key (0~9 digit entry)
TARE key ⇒ Move cursor rightward
NET/GROSS key ⇒ Move cursor leftward
49
to input the setting
UNITS
key. After 5
NET/GROSS
ZSME300000237