Replacement of parts subject to wear and tear

Check the following parts subject to wear and tear and replace where
applicable:
● Therapy light guides (change after 2,000 sterilization cycles or every
two years)
● Silicone key pad of the handpiece (change after 400 treatments/
sterilizations)
● Fiber cutter (change after 400 treatments/sterilizations or every two
years)
● Rechargeable battery (change after 1000 charging cycles or every
two years)
For further information, see chapter "Cleaning, disinfection and
sterilization  [→ 63]".
NOTICE
Spare parts
Only use parts from Dentsply Sirona, see "Spare parts  [→ 25]".