4.4 Initial start-up – procedure for proper assembly
34
CAUTION
Incorrect winding
Incorrect winding of the cable can cause the fiber optics inside the
cable to break.
For cleaning, disinfecting and sterilizing please refer to chapter
"Cleaning, disinfection and sterilization  [→ 63]".
4.4.3
Install remote interlock – optional
Explanation
The interlock is a safety device that stops laser radiation whenever the
door of the treatment room is opened. The interlock circuit must be
connected to a switch that is located near the door of the treatment
room in order to ensure automatic interruption of the laser emission.
NOTICE
The installation must be performed by a qualified electrician who is
also responsible for the installation and maintenance of the electrical
system to which the SiroLaser Blue is connected.
IMPORTANT
Additional or different safety precautions
Additional or different safety precautions required by the applicable
national or local regulations for the protection of dentists, assistant
personnel, or patients must also be observed.
Installation of an interlock with door switch
1. Prepare the interlock plug by connecting the interlock cable with the
interlock plug and by opening the bridge. Please find the technical
data sheet with circuit diagram for the installation of the interlock
circuit in "Appendix C – Safety circuit (interlock)  [→ 82]".
2. Mount the prepared interlock plug into the interlock socket on the
backside of the SiroLaser Blue.
Dentsply Sirona
Instructions for use SiroLaser Blue
65 40 863 D3648
D3648.201.01.07.02    2023-08