Assembly Of Therapy Light Guide - Dentsply Sirona SiroLaser Blue Instructions For Use Manual

Hide thumbs Also See for SiroLaser Blue:
Table of Contents

Advertisement

4 Installation
4.4 Initial start-up – procedure for proper assembly
32
CAUTION
Incorrect winding
Incorrect winding of the cable can cause the fiber optics inside the
cable to break.
For cleaning, disinfecting and sterilizing please refer to chapter
"Cleaning, disinfection and sterilization  [→ 63]".
4.4.2.3

Assembly of therapy light guide

4.4.2.3.1
Area of application
The SiroLaser Blue can be used for additional dental procedures with
two types of reusable therapy light guides of different diameters:
● Light guide (MultiTip 8 mm), diameter: 8 mm
● Light guide (MultiTip 4 mm), diameter: 4 mm
The light guides are delivered non-sterile.
WARNING
After 2,000 sterilization cycles or 2 years which marks the end of the
service period, the MultiTips will have reached their wear limit. Please
check the usage period based on the LOT number (definition of LOT
= week year e.g. 0215 for calendar week 2, 2015). Please replace the
light guide accordingly. The optical output can be reduced.
The MultiTips can be used only with the SiroLaser Blue in the spectral
range of 445 nm -10/+15 nm, 660 nm ± 5 nm and 970 nm -10/+15 nm.
WARNING
If glass rods from other manufacturers are used, physical properties
such as load carrying capacity and transmission behavior may vary.
Dentsply Sirona therefore assumes no liability in such
cases.Therefore, use only Dentsply Sirona glass rods (MultiTip).
Dentsply Sirona
Instructions for use SiroLaser Blue
65 40 863 D3648
D3648.201.01.07.02    2023-08