I-V600
6.4.5. Anomalous situations
If the instrument detects a voltage
of -0.5V≤VPN≤15VDC at the P-N
terminals of the string, it does not
perform
the
test,
emits
a
prolonged
audible
signal
and
displays the message shown in
the screen alongside. Check the
tension on the string
If the instrument detects a voltage
<-0.5VDC at the P-N terminals of
the string, it does not perform the
test, emits a prolonged audible
signal and displays the message
shown in the screen alongside.
Check the connections of the
string poles
If the SOLAR03 remote unit is
active but not connected and not
recording, the instrument does
not perform the test and displays
the message shown in the screen
alongside. Connect the SOLAR03
remote unit
EN - 40