Gravity Acceleration Correction - AND SJ-WP Series Instruction Manual

Compact bench scale
Table of Contents

Advertisement

11.2. Gravity Acceleration Correction

When the scale is first used or has been moved to another location, it should be
adjusted using a sensitivity adjustment weight. But if a sensitivity adjustment weight is
not available, the gravity acceleration correction will compensate the scale. Change the
gravity acceleration value stored in the scale to the value of the area where the scale
will be used. Refer to the gravity acceleration map at the end of this manual.
Note: Gravity acceleration correction is not required when the scale will be
adjusted using a sensitivity adjustment weight at the place where it is to
be used.
1. Refer to
"11.1. Sensitivity Adjustment
sensitivity adjustment mode. The CAL is displayed.
2. Press the TARE switch to enter the gravity acceleration
value setting mode.
3. Change the displayed value using the following switches.
MODE
To shift the blinking digit to the right.
SAMPLE To increase the value of the blinking digit by one.
4. Press the TARE switch. The display shows
returns to the newly stored gravity acceleration value.
5. When sensitivity adjustment using a sensitivity adjustment
weight is to be performed, go to step 3 of
Adjustment Using a
To finish the setting procedure, press and hold the
switch or the CAL switch. The scale returns to the weighing
mode.
Mode" to enter the
Weight".
and
"11.3. Sensitivity
30