Teledyne ASX-560 Installation Manual page 13

Hide thumbs Also See for ASX-560:

Advertisement

5
Guide the probe straight down until the tip of the probe is level with the
bottom of the Z-Drive.
6
Tighten the probe clamps.
7
Use the provided spiral wraps to secure the sample tube to the Z-drive cable.
There should be just a little curve to the free sample tube when the probe is
lowered, and an untangled loop when the probe is raised. Check that the
tubing will not be stretched and will not snag on an obstacle when the probe is
moved to the far corner sample positions.
Loop
8
Optional: Moisten a Kimwipes
remove any aqueous or inorganic contamination from the surface. Then
moisten a new Kimwipes wiper with a chromatography grade solvent such as
n-propanol, isopropanol, or hexane and wipe the probe again to remove any
oil. Let the probe air dry. Be careful not to touch the probe without gloves—
residue from handling the probe can cause the surface tension properties of
the probe surface to change and can lead to material dripping off the probe
during use. Periodic cleaning may be necessary depending on the application.
EAR99 Technology Subject to Restrictions Contained on the Cover Page.
Spiral wrap
wiper with 10% HNO
and wipe the probe to
3
13