Installing Transmitter On Pipe (Remote Type) - Siemens SITRANS FC720 Operating Instructions Manual

Coriolis c
Table of Contents

Advertisement

FC7x0
Installation
3. By using an Allen wrench, remove the bottom fixing screws and rotate the terminal
4. Place the terminal box and tighten the bottom fixing screws using a minimum torque
5. Attach the cover and tighten the fixing screws using a minimum torque of 7.4 Nm.

12.6.4 Installing transmitter on pipe (remote type)

Risk of overheating the transmitter due to increased ambient temperature
WARNING
Failure of measuring electronics
▶ Observe the maximum allowable ambient temperature for the transmitter.
▶ Install the transmitter at a sufficient distance from heat sources. Also note the
Short-circuit hazard caused by penetrating water
WARNING
Failure of measuring electronics
▶ In order to prevent any water from penetrating the flow meter by way of the cable,
Risk of injury and damage to the flow, meter if it is insufficiently attached to
CAUTION
the pipe
▶ Observe the installation notes below.
▶ Tighten screws by using a minimum torque of 7.4 Nm.
102 / 180
box at an angle of 90°, 180° or 270°.
of 7.4 Nm.
temperature of the fixing pipe.
install the transmitter in a way so that the cable gland is not pointed upward.
A5E52748387-AA, 2023/09
Transmitter installation