Revision of Instruction
1995. 23. Mar.
We would like to revise the instruction manual for Nikon Retinomax is revised as follows :
•Add the following paragraph to 2) Automatic charging
"2"
on
Page 18:
Match the right comer of the switch panel of the station to the left side of the
label on the surface of the printer in order to connect the printer to the station
easily.
•Revise
"1)"
clause as follows and add the following
"6)"
clause to
''(6)Advise and confirmation"
on Page 21:
(Wrong)
1) If is normal for the battery to become warm after use or charging.
i
(Correct) 1) It is normal for the battery to become warm after renewing or charging.
(Add )
6) Do not cover the main body, printer, and station with unique covers when charging the
main body battery of printer battery. Charging generates heat, which makes the main
body, printer, and station overheated if the covers are attached. This exerts bad
influence on the batteries, which makes their lives shorter.
•Add the following description to
on Page 24:
In the automatic eye identification mode, the buzzer will beep when an eye to be examined changes to the
other during measurement. Use this mode to check which eye you are measuring.
However, the buzzer may beep repeatedly if the eye identification is not performed properly because the
sensor is in between a right and left eyes.
Note: In this case, the buzzer will beep even when you set the BUZZER option to OFF in the initial setting
(SETUP).
•Add the following description to Page 26:
-[^t@ifetKe^tigm^tsm^tT^iurini^rai|^ -
Values of astigmatism measurement range from -8.0 to +8.0 . However, values ranging from -12.0 to -8.0 and
from +8.0 to +12.0 are also displayed as the reference data.
The measurement values within these extended ranges are displayed with the characters
"C-COVER"
shown under the cylindrical power indication. However, the cylindrical power is printed out without the
"C-COVER"
characters.
•Revise the description in the section "When the measurement is not successfuiiy done
due to instabiiity or quick movement of eyes (QUiCK mode)"
on Page 28 as follows.
(Wrong )
• The QUICK mode increases the speed of measurement due to the automatic fagging being
disabled, but may lose measurement values may generate measured values that have low
accuracy or vary over a wide range. You should not use the QUICK mode for during normal
eye measurement.
i
(Correct)
• The measurement time in the QUICK mode is about half of that in the Normal mode.
However, measurement values may vary over a wide range due to the fogging operation in
the QUICK measurement. You should not use the QUICK mode for during normal eye
measurement.
1