Rough alignment of the patient's eye (see
Figure 6-6).
• Vertical: Align the patient's eye parrallel with
the target line located on the side of the
measuring unit.
• Horizontal; Align the patient's eye parrallel
with the target line located on the top of the
measuring unit.
• working distance: Position the patient's eye
about 40 mm away from Maxs in body.
Placing your other hand as shown in figure 6-7
on the units main body will stabilize the
measurement procedure.
® In the measurement mode, with the patient's eye
roughly aligned, look through the view finder.
When you see the patient's eye on the screen,
position the pupil image towards the center (see
Figure 6-8).
Fine focus the dot by moving the measuring unit
back and forth until the light spot becomes as
small as possible.
When bringing the measuring unit close to the
patient's eye, it will automatically identify the right
and left eyes.
The eye being measured whether right or left
will be
indicated in the screen as shown in the
figure on the right.
(D
When the alignment process is complete, the measurement is automatically started.
Every
time a measurement cycle is done, a beep sounds.
During measurement to reduce
accommodate, the automatic fogging function slightly blurs the target (see page 26).
If you
would like to disable the automatic fogging function, simply press the QUICK key to enter the
Quick mode (see page 28).