Review An Ecg Patient Report - GE MAC VU360 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Recording an ECG or Rhythm
If ...
Hookup Advisor status
is yellow or red, there are
issues with the ECG signal
quality during the recording
of this ECG. The Reject icon
is highlighted in blue to
reflect the status of Hookup
Advisor. The Accept icon is
not highlighted.

Review an ECG Patient Report

When you disconnect patient leads and there is no activity with the MAC VU360
device for 2 minutes, all ECG reports will automatically close.
After the 10 seconds ECG is acquired and the ECG preview is accepted, the patient
report is displayed in the ECG tab in the configured report format for review.
If the CRIT option is enabled on the device, and one or more critical values are
detected during ECG acquisition, a window opens on top of the ECG tab displaying
the notifications for the critical values in the order in which they are detected.
Select Continue to acknowledge the notification and proceed with other tasks.
54
Then ...
Select the Reject icon:
The ECG patient report
is discarded. The ECG
Preview tab closes
and returns to the
Acquisition screen.
  MAC VU360
Resting ECG Analysis System
Next Steps
Start a new ECG on the same
patient. For more information, see
ECG Acquisition Overview on page
47.
 2088531-370-2