These changes apply only to the current patient's test(s). If you start a new
patient, the changes reset to the values configured for the device.

ECG Acquisition Overview

You can record an ECG in pre-acquisition or post-acquisition modes.
Table 15: ECG Acquisition Modes
Acquisition
Mode
Pre-acquisition
Post-acquisition
If the Auto-ECG option is enabled on the device, the device automatically records one
ECG for each new patient. For more information on automatic ECG acquisition, see
Automatically Acquire an ECG on page
To record an ECG patient test manually, see

Automatically Acquire an ECG

Start a test for a new patient. For more information, see
on page
Make sure that your administrator enables the Auto-ECG setting.
The system begins to automatically record an ECG:
• When Auto-ECG setting is enabled.
• If the Hookup Advisor status is green.
• If you are viewing the scrolling waveforms in the Live tab.
The automatic acquisition of ECG happens only once per patient connection.
The Stop Auto ECG icon displays the count of acquisition progress until you record
10 seconds of data. After acquiring 10 seconds of data with good signal quality, the
recording stops and a preview of the ECG patient report displays in the ECG tab.
If you record the ECG before you enter patient information in the Patient Information
screen, you can edit patient information before accepting the preview. For more
information, see
Automatic ECG acquisition is triggered only one time for the current patient test. Start
any new ECG tests for that same patient manually.
2088531-370-2
Description
When you start recording an ECG:
If 10 seconds of ECG data is available, the system records the previous 10
seconds of data for analysis.
If 10 seconds of ECG data is not available, the system continues recording
until 10 seconds of ECG data is obtained.
When you start recording an ECG, the system records the next 10 seconds of
data for analysis.
31.
Accept or Reject an ECG Patient Report on page
  MAC VU360
47.
Record an ECG Manually on page
Resting ECG Analysis System
Recording an ECG or Rhythm
48.
Start a Test for a New Patient
50.
 47