Troubleshooting
Error Message
Cannot perform action while
printing

Report Transmission Errors

Table 33: Errors Encountered During Patient Report Transmission
Error Message
Transmission queue is full. Additional
reports cannot be added to the
queue.
Unable to transmit. Incomplete
patient data.
Unable to transmit one or more
reports. Incomplete patient data.
Report transmission is in progress.
Delete job from queue to edit.
Destination unknown
126
Error Condition
You have tried to perform
below actions while printing a
patient report:
Start a test for a new
patient
Delete the patient report
Change speed, gain, or filter
Access User Menu
Access Orders, Files, or
Queue lists
Power off, lock, log out,
standby or privacy mode
Error Condition
The transmission
queue has reached
its maximum limit
of 1,000 reports.
One or multiple
patient reports
cannot be added
to the transmission
queue because
required fields
in the patient
demographics are
blank or contain
invalid data.
You are trying
to edit a patient
report that is being
transmitted.
The destination is
not found.
  MAC VU360
Resting ECG Analysis System
How to troubleshoot:
Perform only allowed actions.
How to troubleshoot:
Wait for the reports in the queue to
transmit and try again.
1.
Edit the incomplete patient
report(s) to enter missing
patient data.
2.
Retry transmission.
Delete the job from the queue to
continue editing the patient report.
Reconfigure the destination. See
MAC VU360
Resting ECG Analysis
System Setup and Configuration
Manual.
 2088531-370-2