Technical Specifications
Item
Paper size
Paper tray capacity
Table 56: Electrical Specifications
Item
Power supply
Input voltage
Input current range
Input frequency
Table 57: Battery Specifications
Item
Type
Charge time
Battery capacity
Table 58: Other Input Device Specifications
Item
External USB barcode reader
Acquisition Module
162
Description
215 mm x 280 mm (8.5 in x 11 in) (modified letter)
210 mm x 297.5 mm (8.27 in x 11.7 in) (A4)
Holds up to 300 sheets
Description
AC mains or battery operation
100–240 VAC ±10%
1.5A @100V to 625mA @240V
50-60 Hz ± 3 Hz
Description
Replaceable and rechargeable internal battery
Approximately 240 minutes from total discharge when device is off
(standby).
The system shall be able to continuously display scrolling
waveforms without printing for a minimum of six hours when the
maximum total battery capacity is installed and the batteries are
fully charged.
Description
Fixed and variable length types
Symbologies: Code-128, PDF417, Code 39, Interleaved Code 2 of
5, and Data Matrix symbology for characters A-Z (upper case), a-z
(lower case), and 0-9 for all supported languages.
Real-time Hookup Advisor with LED lead quality indicators
Ingress protection level: IPx4
Optional integrated barcode reader