Nehb Electrode Placement - GE MAC VU360 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Table 48: Pediatric Electrode Placement
Item AHA Label
1
V1 red
2
V2 yellow.
3
V3 green.
4
V4 blue
5
V5 orange
6
V6 violet
7
LA black
8
LL red
9
RL green
10
RA white
11
V4R gray
12
V3R gray
13
V7 gray

NEHB Electrode Placement

To acquire a NEHB ECG, use the standard 12–lead electrode placement and items 1, 2
and 3 as shown in the following diagram.
2088531-370-2
IEC Label
Description
C1 red
Fourth intercostal space at the right sternal border.
C2 yellow
Fourth intercostal space at the left sternal border.
C3 green
Midway between location 2 and 4.
C4 brown
Mid-clavicular line in the fifth intercostal space.
C5 black
Anterior axillary line on the same horizontal level as 4.
C6 violet
Mid-axillary line on the same horizontal level as 4 and 5.
L yellow
Left deltoid.
F green
Above the left ankle (alternate placement— upper leg as close
to the torso as possible).
N black
Above the right ankle (alternate placement—upper leg as
close to the torso as possible).
R red
Right deltoid.
C4R gray
Mid-clavicular line in the fifth right intercostal space.
C3R gray
Halfway between 1 and 11.
C7 gray
Same horizontal level of 4 in the left posterior axillary line.
  MAC VU360
Resting ECG Analysis System
Patient Preparation
 147