ECG Specifications

Table 59: ECG Data Acquisition Specifications
Item
Signal input
Dynamic range
Common mode rejection
Input impedance
Patient leakage current
Detection of pacemaker pulse
Pace Annotation
Pace digital sampling rate
Table 60: ECG Data Processing Specifications
Item
ECG Interpretation
Computerized measurements
Additional ECG function
Heart rate meter
ECG data formats
Pre-acquisition
Digital rhythm
Storage
External storage
2088531-370-2
Description
Type CF, Defibrillation-Proof
Defibrillation protection: Per IEC 60601-2-25:2011
AC Differential ± 5mV, DC offset ±300 mV
>130 dB (>100 dB with AC filter disabled)
>10MΩ @ 10 Hz, defibrillator protected
<10 µA
Duration: 0.2 ms to 2.2 ms
Amplitude: 2 mV to 700 mV
Separation: 1 ms or greater
Dedicated pace channel on display and printed reports
75,000 samples/second per channel
Description
Marquette 12SL ECG Analysis Program for Adults and Pediatrics
15-lead analysis (includes measurements of user-selectable
additional 3 leads)
Vectorcardiography
30 to 300 beats per minute (BPM) with an accuracy of ±10% or 5
BPM, whichever is greater. Heart rates outside this range will not
display.
GE Hi-Fidelity ECG, XML
Provides 10 seconds of instantaneous ECG acquisition
Up to 5 minutes of continuous rhythm storage (exportable as a
PDF)
1000 records consisting of 10 second Resting ECG records and
Digital Rhythm records on the device internal memory
USB-compliant flash drive supporting the FAT32 file system
  MAC VU360
Resting ECG Analysis System
Technical Specifications
 163