Report Formats
Item
Name
3
12SL
Interpretation
Statements
4
ECG Header
5
Physician
Information
6
Report Status
7
Waveforms
8
Pace Spikes
9
Page Number
10
Report Format
11
12SL Version
12
Product Model
13
Filter Setting
14
Gain Setting
15
Speed Setting
136
Description
Displays automated 12SL interpretation statements if the report
format is configured to include 12SL interpretation statements.
Clinicians use this information to make decisions about cardiac
care for the patient. The patient report includes ACS interpretation
statements when the ECG is recorded with the ACS option.
The status of Hookup Advisor is based on the 12SL analysis of the
patient report.
Displays the Patient ID, date and time of ECG acquisition in the
configured date and time format, and the name of the institution, if
configured.
Displays the following details:
Referred by: Name of the physician who referred the patient
Secondary ID: Alternate identification number of the patient
Displays the status of the report.
Displays the 10 seconds ECG patient report in the configured report
format for the selected lead set.
Displays pace spikes for patients with a pacemaker. The pace spikes
represent pacemaker pulses.
NOTE:
When two pace spikes are detected so close to each other that
they cannot be uniquely shown on the report, a single pace flag
with two flag tips is printed to indicate this condition.
Displays the page number of the ECG patient report in the format
page x of y, where x is the current page number and y is the total
number of pages.
Displays the configured report format title used to preview the ECG
patient report.
Displays the 12SL version used to analyze the ECG patient report.
Displays the product model.
Displays the filter of the ECG waveform (measured in Hz), with Zero
Phase Distortion (ZPD for High Pass filter).
Displays the gain of the ECG waveform (measured in mm/mV).
Displays the speed of the ECG waveform (measured in mm/s).