Prefilter Installation

MPS Cartridge
Installation
Thermo Scientific
Particulates can damage your membrane, resulting in premature membrane
failure. Therefore, your Thermo Scientific RO system uses a prefilter to
remove particulates from your feed water. Install the prefilter cartridge as
follows:
1. Open the front left door.
2. Remove the prefilter (Part Number D502113) from its packaging and
wet the o-rings with water.
3. Insert the prefilter straight into the left position upper and lower quick
disconnects and push until firmly connected. For easier installation,
connect bottom first.
The MPS cartridge contains a slow dissolving anti-scalant material. The
anti-scalant material combines with water to prevent scale buildup on the
membrane surface. Install the MPS cartridge as follows:
1. Remove the MPS cartridge (Part Number D502114) from its
packaging and wet the o-rings with water.
2. Insert the MPS straight into the middle position upper and lower
quick disconnects and push until firmly connected. For easier
installation, connect bottom first.
Note Figure 4-5 shows the correct
sequence of prefilter, MPS and
carbon filter. Cartridges must be
installed in correct order to
operate properly.
s
Note It is important to replace the
MPS cartridge once the material is
half dissolved.
s
Unpacking and Installation
Prefilter
MPS cartridge
Figure 4-5. Cartridge Installation Order
Reverse Osmosis System
Section 4
Carbon filter
4-7