GE Definium 5000 Operator's Manual page 18

X-ray system
Table of Contents

Advertisement

WARNING:
For diagnostic X-ray equipment specified to be used in combination with accessories or
other items not forming part of the same; attention to the possible adverse effect arising
from materials located in the X-ray beam. Refer to Table 2-1 for maximum attenuation
equivalent of possible materials located in the x-ray beam.
Table 2-1 Maximum Attenuation Equivalent mm AL
Collimator without spectral filters
Collimator spectral filtration of 0.1mm Cu
Collimator spectral filtration of 0.2mm Cu
Detector Housing at 100 kV
Image Pasting patient barrier
Non-elevating table
Non-elevating table with carbon-fiber table top
Elevating table with carbon-fiber table top
Definium 5000 X-Ray System
Operator Manual
Item
5197809-1EN
Rev. 4 (13 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter 2: Safety and Regulatory
mm Al
2
3.1
4.5
.9
<0.08
<1.05
<1.0
<1.2
2-4