The registration markers (brackets) show the general area of pasted overlap (Figure A-17). Use the Image
Magnifying Glass function to review registration quality.
Figure A-17 Close-up of registration marker and mis-registered area
Follow this process to paste images.
1. From the tool selection list, select Reprocessing.
2. Press the [PASTE] button.
NOTE: It may take a few moments for images to appear.
3. Review the pasted image.
If the image is satisfactory, press [APPLY TO IMAGE] to save the composite and return to the main
Image Viewer screen.
If the pasted image is not satisfactory, follow the steps described in
make adjustments.
NOTE: The pasted composite will not be saved until [APPLY TO IMAGE] is pressed.

Correct Pasting

Follow this process to adjust a pasted image.
1. Select a region to re-paste from the drop-down list (Figure A-18). For example: 1/2, 2/2.
[RESET REGION] removes any adjustment from the selected region.
Figure A-18 Select region drop-down list.
2. Press and drag a red cross-hair to move it (Figure A-19).
Definium 5000 X-Ray System
Operator Manual
5197809-1EN
Rev. 4 (13 February 2008)
© 2008 General Electric Company. All rights reserved.
Appendix A: Image Pasting
Correct Pasting
(p. A-10) to
A-10