Emerson Rosemount 8600D Series Reference Manual page 89

Vortex flowmeter
Hide thumbs Also See for Rosemount 8600D Series:
Table of Contents

Advertisement

Reference Manual
00809-0100-4860, Rev BC
Note
Be sure to fully check all other troubleshooting possibilities before removing the sensor.
Also, please note that the sensor is a complete assembly and cannot be further disassembled.
Tools needed
1.
2.
Removable bracket
3.
Figure 5-8. Removable Bracket Assembly
4.
5.
6.
Troubleshooting
/
-in. (4 mm) hex wrench
5
32
5
/
-in. (8 mm) open end wrench
16
5 mm hex wrench
Suction or compressed air device
Small, soft bristle brush
Cotton swabs
Appropriate cleaning liquid: water or cleaning agent
De-pressurize the flow line.
Remove the electronics housing (see
Loosen the four bracket anchor bolts with a 5 mm hex wrench (See
Assembly Figure
5-8).
Bracket Anchor Bolts
Sensor Bolts
Remove the bracket.
Loosen Sensor bolts with 5 mm hex wrench.
Remove sensor bolts, sensor, and gasket.
Get other manuals https://www.bkmanuals.com
Section 5: Troubleshooting
"Replacing the electronics housing" on page
January 2013
79).
"Removable Bracket