Display
The warning message 'Device in critical tilt' serves as preventive information to avoid
malfunction caused by the device tilting over (for example, a blocked bacterial and viral filter in
the secretion canister).
The warning message function is activated in the factory settings. If the warning message
'Device in critical tilt' is not desired, it can be deactivated (see chapter '4.9 User settings' on
page 43).
Generally, it is recommended to activate the warning message function 'Device in critical tilt' to
avoid malfunction caused by the device tipping over.
WARNING
Device temperature too high
Provide sufficient ventilation!
Check fan!
48
Cause of error
If the device is in a tilted
position, the warning message
'Device in critical tilt' appears.
If the therapy has not been
started after new installation/
start-up of the device, the
warning message 'Inactive
therapy' appears without tone.
If the therapy has already
started and has been
interrupted by the pause
symbol, the warning message
'Inactive therapy' appears with
tone.
The device must no longer be
operated.
Possible causes:
• Battery or
• pump defective.
Device temperature too high.
• Device is in the sun or near a
heater.
• Ventilation slots are covered.
• Fan is defective.
Troubleshooting
Place the device in an upright
position. The warning
message is automatically
deactivated.
Touch the play symbol to
start the therapy.
If the warning message is
suppressed, it reappears
after one minute, but without
tone.
Touch the play symbol to
start the therapy.
If the warning message is
suppressed, it appears again
with tone after one minute.
Contact ATMOS Service.
• Place the device in a cooler
location.
• Please ensure sufficient air
ventilation.
• Please contact ATMOS
Service.