5.0 Warning messages

In the event of a warning message, the key lock is automatically unlocked!
In the event of a warning message, the system automatically switches to the warning message
window. A warning message is displayed. This contains advice for the elimination of the cause of
error. The acoustic warning message is triggered at the same time.
Display
Cause of error
If the target vacuum is not
reached, the alternating
warning message 'Vacuum too
low' and 'Secretion canister full
or hose blocked' is activated.
Possible reasons for this
warning message are:
Leakages
If the target vacuum is not
reached, the alternating
warning message 'Vacuum too
low' and 'Secretion canister full
or hose blocked' is activated.
Possible reasons for this
warning message are:
• Blockages
The measurement of an
excessively high vacuum results
in the display showing the
warning message 'Vacuum too
high'.
Possible reasons for this
warning message are:
• Ventilation valve is defective.
• There are further vacuum
sources in the drainage area.
If the voltage of the battery
falls below a specific value, the
warning message for 'Battery
low' is displayed.
Troubleshooting
• Check for leakage:
- Connection from the
hose system to the
patient catheter
- Connection from the
hose system to the
secretion canister
- Secretion canister
connection
• Contact ATMOS Service
• Check for blockages:
- Secretion canister
- Hose
- Filter in the secretion
canister
- Filter in the measuring
hose
• Contact ATMOS Service
• Remove vacuum sources
• Contact ATMOS Service
Connect device to the supply
network. The battery is
charged and the state of
charge is indicated in the
display.