2.3 Warning

WARNING
WARNING
WARNING
WARNING
WARNING
Electric shock due to damaged connecting cables
Prior to use check the device and connecting cables for damages.
Damaged cables must be replaced!
Electric shock due to voltage!
Prior to disconnecting the battery from the mains pull the plug from
the electrical socket, after that you can disconnect the connecting
cable from the battery.
Electric shock due to voltage!
Disinfectant or secretion should not enter the device or battery.
If disinfectant/secretion has penetrated the device, then it must be
dried thoroughly and subsequently an effi ciency control must be
conducted.It should be tested if the target vacuum can be reached in a
closed system, as well as if there is a fl ow > 4 l/min after a short while,
when the system is open. If not, the device must undergo a service
check before it resumes operation.
The mains adapter should not get wet.
Don´t take a shower / bath with the device.
Danger of explosion due to unobserved ambient conditions!
The ATMOS
C 051 Thorax may only be operated in rooms used for
®
medical purposes, but not in areas subject to explosion hazards
or in oxygen rich environments.
Explosion hazardous areas may result from the use of fl ammable anae-
stetics, skin cleansing agents and skin disinfectants.
Pay attention to the ambient conditions specifi ed in chapter 11.0.
Risk of infection
This product is not re-sterilizable. Repeated reuse of components
which are marked with
these components lose their function and there is a high infection
risk.
Repeated reuse of canister and hose systems can lead to infections.
Canister and hose system should only be used once on every patient.
For hygienic reasons we recommend an exchange of both canister and
hose system at the same time.