LAPAROSCOPIC OPERATION

FOR BEST RESULTS:
1. Installation of Crystal Vision should already be
completed according to the installation instructions.
Connect SAFEGUARD BLUE
Filter with Built-in Fluid Trap to CRYSTAL VISION
front panel. Connect sensor LASER SENSOR and/or the
ESU/RF/Universal SENSOR, as desired, to CRYSTAL
®
VISION
front panel.
2. Be familiar with all operating controls as described in
Description of Switches, Control Buttons, and Indicators.
SET-UP when using ESU in Laparoscopic Procedures:
1. Install and connect a 10mm Laparoscope CANULA
(TROCAR) to the insufflator.
2. Install and connect the I/A tubing to Crystal Vision on
one end, and other end to the CANULA (TROCAR).
3. Install a second 10mm CANULA (TROCAR) to use
with 5mm ESU instrument.
4. Upon installation the user may detach the laparoscopic
tubing from Laparoscope/Canula and attach to 10mm
CANULA (TROCAR) used for ESU instrument and
install SMOKE EVACUATOR TUBING to the 10 mm
Laparoscope CANULA (TROCAR).
5. The insufflator FLOW should be set to Maximum, and
the pressure should be set to surgeon's needs. Refer to
Figure 19.
At the end of procedure remove all instruments and
CANULA except the one with Smoke Evacuator
attached. Set the Smoke Evacuator Flow at maximum
in LAP mode, time at 30 seconds and press the manual
button-the entire quantity of CO
filtered and removed safety so no gas will escape in OR,
protecting the surgeon and OR stuff from bacteria and
viruses.
SET-UP when using CO
Procedures:
1. Install and connect the Laser Laparoscope luer lock to
the insufflator.
2. Install and connect the I/A tubing, to Crystal Vision on
one end, and to the CANULA (TROCAR) on the other
end. Refer to Figure 20.
At the end of procedure remove all trocars and
instruments except the one with Smoke Evacuator
attached.
Set the Smoke Evacuator Flow at maximum in LAP
mode, time at 30 seconds and press manual button-in few
seconds the abdomen will be deflated, the gas will pass
through the SAFEGUARD BLUE
®
Hydrophobic ULPA
®
from the abdomen is
2
Laser in Laparoscopic
2
®
Hydrophobic ULPA
Filter with Built-in Fluid Trap, filtered, and no bacteria
and viruses will escape in OR.
Figure 19: Laparoscopy Set up when using ESU and Crystal
Vision
Figure 20: Laparoscopy Set up when using CO
Crystal Vision