8. FLOW SETTING control buttons (Figure 16) sets the
maximum flow that is desired by the surgeon. This value
is indicated on the FLOW SETTING display.
9. FLOW READING (Figure 16) is registering the amount
of gas and vapor that is actually flowing at the present
time. This value should be zero, when the vacuum pump
is not turned on.
10. CHANGE FILTER indicator is located above filter (Fig.
17). It illuminates when there is a reduced flow into the
CRYSTAL VISION
®
.
Hydrophobic ULPA Filter with Built-in Fluid Trap
should be changed when this light first illuminates. Do
not attempt to clean, or re-use the SAFEGUARD
®
BLUE
Hydrophobic ULPA Filter with Built-in Fluid
Trap.
Dispose
of
Hydrophobic ULPA Filter with Built-in Fluid Trap
according to your institution's biological waste protocol.
This indicator will also come ON when there is a total
occlusion. If the OCCLUSION indicator is also lit, be
sure to clear the obstruction that caused it first and then
check the CHANGE FILTER indicator. If it is still
illuminated and the OCCLUSION indicator is not, then
the SAFEGUARD BLUE
with Built-in Fluid Trap needs to be changed.
11. OCCLUSION indicates that flow into the CRYSTAL
®
VISION
has stopped (Figure 17). The CHANGE
FILTER light will also come ON at this time. The
operator should check for kinked tubing, stopcocks that
are turned OFF, clogged insufflator, use by date of Large
Coconut Charcoal Output Filter, or a completely clogged
SAFEGUARD BLUE
Built-in Fluid Trap. In order to prevent pump failure, the
®
CRYSTAL VISION
lamp is illuminated.
12. OVER PRESSURE indicates pressure exceeding 27
mmHg in the pneumoperitoneum (Figure 17). At
pressures over 30 mmHg, the vacuum pump starts in order
to reduce the pressure in the pneumoperitoneum. This
situation can be caused by many factors. Frequently, the
surgeon presses on the abdomen and this causes intra-
abdominal pressure to rise. Correct the cause before
continuing.
13. The "NO PATIENT" indicator (Figure 17) turns on when
the CRYSTAL VISION® is in the LAP MODE, when
zero pressure, or atmospheric pressure, is present at the
SAFEGUARD BLUE
Built-in Fluid Trap. The tubing set may not be connected
to the Trocar sheath or to the SAFEGUARD BLUE
Hydrophobic ULPA Filter with Built-in Fluid Trap; the
valve on the Trocar sheath may be turned off; or the
SAFEGUARD BLUE
Built-in Fluid Trap may not be attached to the
SAFEGUARD BLUE
Built-in Fluid Trap connector.
The SAFEGUARD BLUE
the
SAFEGUARD
BLUE
®
Hydrophobic ULPA Filter
®
Hydrophobic ULPA Filter with
pump will not activate when this
®
Hydrophobic ULPA Filter with
®
Hydrophobic ULPA Filter with
®
Hydrophobic ULPA Filter with
®
®

Figure 16: Selecting the desired flow.

Figure 17: Front Panel Indicators.

®