This product should only be powered as described in the manual. To prevent equipment damage, select the proper
voltage outlet.
7.
To prevent damage to the equipment, read the instructions in the equipment manual for proper input voltage.
8. Keep unit at operating environment for at least 6(six) hours before use, if unit was exposed to extreme shipping
and storage conditions.
9. Make sure the unit is in a safe and stable environment as to prevent falling or being dropped, which may cause
damage.

COMPATIBILITY

Refer to below compatibility information, Specifications Section, and Installation/Operations Instructions to confirm
that this Crystal Vision Model is compatible with the accessories being used.
The Smoke Evacuator was tested electrically to meet the requirements ANSI/AAMI ES 60601-1 Medical electrical
equipment— Part 1: General requirements for basic safety and essential performance. This unit is compatible with
other IEC 60601-1 certified units.
Electrosurgical equipment (ESU) connected to the auxiliary Mains outlet must be certified according to IEC60601-1,
including Medical Electrical System aspects. Everybody who connects additional equipment to the auxiliary mains
outlet configures a medical system and is therefore responsible that the system complies with the requirements of IEC
60601-1.
For the electromagnetic compatibility (EMC), this smoke evacuator complies with EMC standard for medical
electrical equipment (IEC 60601-1-2).
List of compatible I.C. Medical products:
1. ESU Sensor
2. RF Sensor, Shielded
3. UNIVERSAL BLUE™ RF Sensor
4. Laser Sensor
5. Foot Switch
6. SAFEGUARD BLUE
7. Large Coconut Charcoal Output Filter
8. Intra-Abdominal Plume Eliminator Tubing Set
9. Speculum Tubing
10. Smoke Evacuator Wand
11. Smoke Evacuator Tube
12. ESU Shroud
13. PenEvac1
14. Non-Telescopic PenEvac
15. The power cords for smoke evacuator units should be grounded, medical-grade type
In addition, new products released after the introduction of this product may also become compatible with this Crystal
Vision Model. For further details, contact I.C. Medical.
WARNING
If combinations of equipment other than those shown within this manual are used, the full responsibility is assumed
by the medical facility.
Connecting additional equipment, other than one found compatible with, to the auxiliary mains outlet or other inputs
(ESU, Laser) will increase chassis leakage.
Using incompatible equipment can result in patient injury and/or equipment damage.
®
Hydrophobic ULPA Filter with Built-in Fluid Trap