ESU, RF, or UNIVERSAL SENSOR
INSTALLATION:
If you are going to use your CRYSTAL VISION
electrosurgical unit (ESU) monopolar or bipolar, ultrasonic
device, harmonic scalpel, proceed with the following:
RF SENSOR
For use with all Crystal Vision
INSTALLATION/OPERATION:
RF Sensor when used with Monopolar Devices
(Figure 2; Option 2; Placement 1 or 2):
1. Plug the RF Sensor's connector into Crystal Vision Smoke
Evacuator's ESU/Laser port.
2. Place the RF Sensor, on the monopolar device's cable (e.g.
PenEvac), by running it through the sensor's wire clip.
3. Make sure the sensor is installed on top of the wire, with
the clip side facing down, close to the monopolar device's
plug, to prevent capture of residual RF signal from
surrounding devices.
4. Plug the monopolar device (e.g. PenEvac), into the
Monopolar port of the ESU Generator.
5. Select the "Monopolar" option on ESU Generator.
6. Set ESU Generator's Cut and Coag value.
7. Activate the monopolar device by depressing cut or coag
button. When the monopolar device activates, the RF
Sensor should turn ON the Crystal Vision Smoke
Evacuator, automatically.
8. When not in use, put away the RF Sensor by attaching the
Velcro side of the RF Sensor to the smoke evacuator.
Perform first-time installation by peeling the back film
from the Velcro loop tape, pressing it firmly to a clean and
dry area on the side of the Smoke Evacuator.
To reuse the RF Sensor simply pull it off from the Smoke
Evacuator's side.
RF Sensor when used with Bipolar Devices
(Figure 2; Option 1; Placement 1 or 2):
Plug the RF Sensor's connector into Crystal Vision
1.
Smoke Evacuator's ESU/Laser port.
Place the RF Sensor, on the bipolar device's cable, by
2.
running it through the sensor's wire clip.
Make sure the sensor is installed on top of the wire, with
3.
the clip side facing down, close to the bipolar device's
plug, to prevent capture of residual RF signal from
surrounding devices.
Plug the bipolar device into the Bipolar Port of the ESU
4.
Generator.
Select the "Bipolar" option on ESU Generator.
5.
Activate the bipolar device. When the bipolar device
6.
activates, the RF Sensor should turn ON the Crystal
Vision Smoke Evacuator, automatically.
®
with an
®
Models.
Figure 2: ESU and RF Sensor installation when used
with an electrosurgical generator (ESU)
5