Set the FLOW SETTING for a flow slightly LESS than the
flow that the insufflator can provide.
Push the MANUAL button.
Observe the effect of a complete cycle by noting the
pneumoperitoneal pressure on insufflator gauge/indicator
before the CRYSTAL VISION
immediately after the pump stops. The intra- abdominal
pressure should not drop more than 1-2 mmHg. If it does,
reduce the flow on the CRYSTAL VISION
SETTING and repeat the process until the pressure drops
very slightly. It is important to set the CRYSTAL VISION
flow as high as possible without diminishing the
pneumoperitoneal pressure.
The FLOW SETTING indicates the desired
maximum Flow.
display indicates the actual Flow occurring
at that moment. The meter will read zero
when the pump is off.
ORANGE in OPEN Mode and GREEN in
LAP Mode.
3. The MANUAL button can be used to start the vacuum
pump if smoke still exists when the TIME SETTING is
set for the maximum amount of time. Be sure to monitor
the pneumoperitoneal pressure so that it does not fall more
than a few mmHg (adjust the FLOW SETTING, if
necessary).
OPERATIONAL HINTS:
The TIME SETTING can be used to
conserve insufflator gas by reducing the
amount of time that the Crystal Vision
pumps after lasing stops.
requires
more
circulating nurse, or the laser operator.
4. CHANGE FILTER illuminates when the SAFEGUARD
®
BLUE
Hydrophobic ULPA Filter with Built-in Fluid
Trap needs to be changed.
®
starts to operate and
®
FLOW
The FLOW READING
The meters are
This process
attention
from
5. OCCLUSION indicator lights when flow virtually ceases
from the pneumoperitoneum. Check for kinked tubing,
closed Luer lock on the Trocar sheath, severely clogged
SAFEGUARD BLUE
Built-in Fluid Trap, heavy fluid accumulation in the fluid
trap, or obstructions in the Trocar sheath. The CHANGE
FILTER indicator always lights when the OCCLUSION
indicator comes on.
6. OVER PRESSURE occurs when a pressure more than 27
mmHg (±10% and 1 digit) is detected by the CRYSTAL
®
VISION. When this happens, the OVER PRESSURE
lamp and audio indicators turn ON. If the pressure
exceeds 30 mmHg (±10% and 1 digit) the CRYSTAL
VISION pump starts and attempts to reduce the over
pressure situation.
PRESSURE!
7. NO PRESSURE indicator illuminates when the Trocar
sheath is turned off, the tubing set is not connected to the
Trocar
sheath,
Hydrophobic ULPA Filter with Built-in Fluid Trap is not
connected to the SAFEGUARD BLUE
ULPA Filter with Built-in Fluid Trap connector.
1. Luer lock on the cannula is open,
2. Tubing is not crimped,
3. SAFEGUARD BLUE
Built-in Fluid Trap is clean, may need to be replaced if it
is clogged,
4. Inspect for leaks at the Trocar sheaths, tubing, or
instruments,
5. Increase the TIME SETTING if there are no leaks.
the
6. If TIME SETTING is set to maximum time, use the
MANUAL button until the abdominal cavity is smoke
free. Do not use it excessively, or the pneumoperitoneum
will collapse.
.
17
®
Hydrophobic ULPA Filter with
DO NOT EXCEED 30 mmHg
or
the
SAFEGUARD
IF SMOKE PERSISTS WHEN
THE PUMP STOPS BE SURE
THAT:
®
Hydrophobic ULPA Filter with
®
BLUE
®
Hydrophobic