LASER SENSOR INSTALLATION

If you are going to use your CRYSTAL VISION® with a
laser, proceed with the following:
1. Attach the Sensor Cable (Figure 6) to the LASER
connector on the CRYSTAL VISION® and to the
Sensor Connector Box.
2. The Sensor Assembly has three parts: the Sensor
Connector Box, the Transmit Sensor, and the Receive
Sensor. Both Sensors have double-backed tape on one
side and an infrared lens on the side opposite the tape.
The Receive Sensor has a red indicator lamp that will
light when the sensor assembly is plugged into the
operating CRYSTAL VISION
(NOTE: When the red light goes off, CRYSTAL
®
VISION
will start to operate).
Test the Sensor Assembly for proper operation:
a)
Plug the Sensor Cable into the Sensor
Connector Box and the LASER connector of the
CRYSTAL VISION®.
b)
Plug the CRYSTAL VISION® into an AC
outlet and turn on the Power Switch on the Control
Box back panel and front panel.
c)
Align the Transmit Sensor Lens and the
Receive Sensor Lens until the red light goes off and
the CRYSTAL VISION® starts.
d)
Move the Sensor until the red light comes on.
The pump in the CRYSTAL VISION
operating sometime within 30 seconds after the light
goes out. (NOTE: the actual amount of time that is
required for the pump to cease operation is
determined by the TIME adjustment on the
CRYSTAL VISION
®
3. Place the foot switch for the laser on an easily accessible
work surface (Figures 7 & 8).
4. Inside the foot switch housing (Figure 8), position the
Transmit Sensor and the Receive Sensor on opposite sides
of the foot switch. Do not position the sensors on the sides
of the foot switch.
WITHOUT
REMOVING
COVERING ON THE TAPE, position them so that the
red light comes on.
®
.
®
will stop
front panel).
THE
PROTECTIVE

Figure 6: Laser Sensor Installation

Figure 7: Laser Sensor Installation to the SIDE of the
laser foot switch HOUSING.
8

Advertisement

Table of Contents
loading

Related Products for IC Medical CRYSTAL VISION 450D

This manual is also suitable for:

Icm-450-0000

Table of Contents