Chapter 3 – Initial operation
Chapter 3 – Initial operation
Information contained in this chapter:
Who is allowed to set up, install and operate the HD 450/HD 450 Injection
Other requirements that must be fulfilled for the setting up and installation
Different ways of setting up the unit
How to top up regenerative salt
Filling processing media
Filling rinsing agent
Turning the device on and off
16
Requirements for setting up, installing and
operating the unit
The HD 450/HD 450 Injection is to be set up, installed and
operated only by persons who have proof of their competency or
are operators of a practice.
Observe all instructions in this chapter when operating the unit
for the first time.
According to current VDE regulations, the HD 450/HD 450 Injection
is not suitable for use in areas that have an explosion risk.
The HD 450/HD 450 Injection is intended for use away from patient
areas. The minimum distance from the treatment location must be
at least 1.5 meters in radius.
Should these instructions not be followed, damage to the
HD 450/HD 450 Injection and/or injury may result.

Requirements for the location of the unit

Location The HD 450/HD 450 Injection is to be set up in a dry and dust-free
location. The humidity at the unit's location is to be 30 - 60 % and the
ambient temperature 15 - 35°C. Should this not be the case, this may
cause malfunctions in the unit. Please ensure that the unit is on a level
surface.
However the unit is set up, there are to be no installations in the area
near the rear panel.
When setting up or installing the HD 450/HD 450 Injection, ensure
that adjacent built-in cabinets or other pieces of furniture are resistant
to condensed water.
Should this not be the case, any water vapor and/or steam
emitted may cause damage to surrounding installations.

Connections

The electrical connections and the connections for water supply
and waste water are to be installed only by authorized personnel.
Failing to do so may result in short circuits, fire, water damage
and/or electrical shocks. Serious injury may result.