Trouble Shooting Guide - IC Medical CRYSTAL VISION 450D Operating And Installation Manual

Table of Contents

Advertisement

SYMPTOM
Be sure power cord is plugged in.
CRYSTAL
Check power switch on back panel to see if it is turned on.
VISION
®
Blown Fuse. (Have technically qualified personnel replace as follows: Remove fuse cover with small screw driver,
Won't Turn ON
remove fuse holder, replace fuse with F4AH 250V fuse; replace fuse holder and replace fuse cover).
Check Outlet for voltage reset circuit breaker if needed.
If none of the suggestions resolves this problem, then Unit is malfunctioning, please have technically qualified
personnel check and correct or contact I. C. Medical, Inc.
Occlusion Indicator is on. See solutions under
Pump Won't Start
Press the manual button, if pump starts then check the following;
Activation Sensor assembly is not installed, or it is not installed correctly. If
red light is "on" and that it goes out as soon as the foot pedal is depressed. If Electrosurgical Sensor is used, see
"ESU SENSOR
Sensor cable not connected to CRYSTAL VISION
used).
Press the manual button, if pump does not start then Unit is malfunctioning, please have technically qualified
personnel check and correct or contact I. C. Medical, Inc.
Unplug Sensor from the Unit, adjust TIME SETTING to minimum. If after 2 seconds pump turns off then the
Pump Runs
following may apply;
Continuously
Laser Sensor assembly not installed correctly. Be sure red light on the sensor is "ON" when the pedal is NOT
depressed and that it goes out when the pedal is depressed.
Debris blocking the light beam at the sensor assembly.
ESU Sensor is defective and will need to be repaired.
Unplug Sensor from the Unit, adjust TIME SETTING to minimum. If after 2 seconds pump does not turn off
then Unit is malfunctioning, please have technically qualified personnel check and correct or contact I. C. Medical,
Inc.
For Laparoscopic procedures, Gas is not allowed to flow freely from the pneumoperitoneum to the CRYSTAL
OCCLUSION
VISION
Indicator Is ON
clogged SAFEGUARD BLUE
continuing.
For open procedures, Blockage is occurring during the procedure, determine the blockage and clear away.
Check calibration of Occlusion Indicator (LAP flow & OPEN flow). See
Instructions.
If none of the suggestions resolves this problem, then Unit is malfunctioning, please have technically qualified
personnel check and correct or contact I. C. Medical, Inc.
Pressure exceeds 27 mmHg (±10% and 1 digit) in the pneumoperitoneum. Insufflator or argon beam
OVER PRESSURE
coagulator supplying too much gas. External Pressure (i.e., from physician's arm, etc.) to the pneumoperitoneum
Indicator Is ON
is raising the internal pressure. Correct the problem before continuing.
Check calibration of Over Pressure Indicator. Also check adjustment of the Over Pressure Threshold to Start
Pump. See
If none of the suggestions resolves this problem, then Unit is malfunctioning, please have technically qualified
personnel check and correct or contact I. C. Medical, Inc.
The CRYSTAL VISION
NO PATIENT
BLUE
Indicator Is ON
on the Trocar sheath is open.
Pneumoperitoneum may be at room pressure (no action required).
Check calibration of No Patient Indicator. See
If none of the suggestions resolves this problem, then Unit is malfunctioning, please have technically qualified
personnel check and correct or contact I. C. Medical, Inc.

TROUBLE SHOOTING GUIDE

INSTALLATION."
®
. Check Trocar sheath stopcock, pinched tubing, tissue/fluids blocking the Trocar channel, or a severely
®
Hydrophobic ULPA Filter with Built-in Fluid Trap. Clear obstruction before
Hospital Level Calibration
®
is not sensing pneumoperitoneal pressure. Be sure tubing set and SAFEGUARD
®
Hydrophobic ULPA Filter with Built-in Fluid Trap are properly connected. Also, be sure the stopcock
PROBLEM/RESOLUTION:
"Occlusion Indicator."
®
front panel (and connected to foot switch if Laser Sensor is
Instructions.
Hospital Level Calibration
23
Laser Sensor
is used, be sure the
Hospital Level Calibration
Instructions.