Figure 3: Universal Blue Rf Sensor Installation When Used With An Electrosurgical Generator (Esu) - IC Medical CRYSTAL VISION 450D Operating And Installation Manual

Table of Contents

Advertisement

When not in use, put away the RF Sensor by attaching
7.
the Velcro side of the RF Sensor to the smoke
evacuator. Perform first-time installation by peeling the
back film from the Velcro loop tape, pressing it firmly
to a clean and dry area on the side of the Smoke
Evacuator.
To reuse the RF Sensor simply pull it off from the Smoke
Evacuator's side.
UNIVERSAL BLUE RF SENSOR
For use with all Crystal Vision® Models
INSTALLATION/OPERATION:
UNIVERSAL BLUE™ RF Sensor when used with
Monopolar Devices
(Figure 3; Option 2; Placement 1 or 2):
1. Plug the UNIVERSAL BLUE™ RF Sensor's connector
into Crystal Vision Smoke Evacuator's ESU/Laser port.
2. Place the UNIVERSAL BLUE™ RF Sensor, on the
monopolar device's cable (e.g. PenEvac), making sure the
sensor is installed on top of the cable, facing down, to
prevent capture of residual RF signal from surrounding
devices.
3. Use both of the self-adhering Velcro straps to wrap around
the sensor and over the monopolar device's cable, to secure
the cable to the sensor at both ends of the sensor.
4. Plug the monopolar device (e.g. PenEvac), into the
Monopolar port of the ESU Generator.
5. Select the "Monopolar" option on ESU Generator.
6. Set ESU Generator's Cut and Coag value.
7. Activate the monopolar device by depressing cut or coag
button. When the monopolar device activates, the RF
Sensor should turn ON the Crystal Vision Smoke
Evacuator, automatically.
UNIVERSAL BLUE™ RF Sensor when used with
Bipolar Devices
(Figure 3; Option 1; Placement 1 or 2):
1. Plug the UNIVERSAL BLUE™ RF Sensor's connector
into Crystal Vision Smoke Evacuator's ESU/Laser port.
2. Place the UNIVERSAL BLUE™ RF Sensor, on the
bipolar device's cable, making sure the sensor is installed
on top of the cable, facing down, to prevent capture of
residual RF signal from surrounding devices.
3. Use both of the self-adhering Velcro straps to wrap
around the sensor and over the bipolar device's cable, to
secure the cable to the sensor at both ends of the sensor.
4. Plug the bipolar device into the ESU Generator.
5. Select the "Bipolar" option on ESU Generator.
6. Activate the bipolar device. When the bipolar device
activates, the RF Sensor should turn ON the Crystal
Vision Smoke Evacuator, automatically.