The CHANGE FILTER light should
always
OCCLUSION LAMP.
4. Repeat for medium and maximum flows in the OPEN
MODE.
Adjustment:
Only make adjustments when necessary.
1. Set up unit according to SETUP FOR SECTION 4.
2. Activate the unit.
3. Adjust VR8 clock-wise (to increase sensitivity), until
the CHANGE FILTER lamp just comes on.
4. Adjust VR8 counter clock-wise (to decrease
sensitivity), until the CHANGE FILTER lamp just
turns off.
Adjust until the CHANGE FILTER lamp works
throughout the flow range
OCCLUSION: OPEN MODE (maximum):
Test:
1. Set up unit according to SETUP FOR SECTION 4.
2. Now activate the unit.
3. Completely occlude the smoke tube with your finger.
4. The OCCLUSION lamp comes on.
5. Check the OCCLUSION at minimum, medium, and
maximum flows.
Adjustment:
Only make adjustments when necessary.
1. Set up unit according to SETUP FOR SECTION 4.
2. Activate the unit.
3. Completely occlude the smoke tube with your finger.
4. Adjust VR7 until the OCCLUSION lamp comes on.
5. Adjust VR7 clock-wise to increase sensitivity or
counter clock-wise to decrease sensitivity.
6. Adjust VR7 until the OCCLUSION lamp works
throughout the flow range.
For
emergency
occlusion occurs prematurely turn VR6
for LAP MODE or VR7 for OPEN MODE
counter clock-wise ½ turn. This should
solve the problem.
I. C. Medical, Inc. personnel or those individuals, who
have been fully trained by I. C. Medical, Inc., should only
make any other adjustments.
come
on
before
procedures,
the
when