CRYSTAL VISION
Model 450D is intended for use in the electromagnetic environment specified below. The user of the CRYSTAL VISION
Model 450D should assure that the CRYSTAL VISION
IEC 60601-1-2 Requirements:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
®
Model 450D is used in such environment.
Compliance
Electromagnetic environment - guidance
Level
Group 1
The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class A
The equipment is suitable for use in all establishments other than
domestic, and may be used in domestic establishments and those
Class A
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes,
Full compliance
provided the following warning is heeded: Warning: This
equipment/system is intended for use by healthcare professionals
only. This equipment/ system may cause radio interference or may
disrupt the operation of nearby equipment. It may be necessary to
take mitigation measures, such as re-orienting or relocating the
equipment or shielding the location
30
®